Greisinger GGA581 Atmospheric Oxygen Sensor

Greisinger GGA581 Atmospheric Oxygen Sensor. Greisinger, GGA581, Greisinger GGA581, Atmospheric, Greisinger Atmospheric, GGA581 Atmospheric
يمكنك الحصول على عرض الآن: info@industrialmaterialcenter.com
رقم المنتج GGA581
المصنع Greisinger
Artikelnummer 610031

Description: GGA581 ( Greisinger )

Greisinger GGA581 Atmospheric Oxygen Sensor



Product Name: Greisinger GGA581 Atmospheric Oxygen Sensor
Product Brand: Greisinger
Product Code: GGA581
Product Artikel: 610031
Atmospheric oxygen sensors, closed sensor type with pressure connection For low oxygen concentrations, with pressure connection, suitable for GMH 569x GENERAL: For devices with external pressure port (GMH 5695/3695) this housing is optimal.  Especially for systems with high or low pressure or with existing back pressure by flow. APPLICATION: It can be screwed airtight (Attention: Observe permissible operating pressure!).  The device-pressure port is connected to the sensor pressure port. The device measures and compensates for the actual pressure at the sensor. SPECIFICATIONS: Sensor element:  GOEL 381 Oxygen-partial pressure probe, mounted in external sensor housing replaceable (temperature sensor mounted in housing) Specific features:  for the lowest O  2   concentrations, for protective gases, in general, precise and very small measurements and above 35 vol .% O  2 Measuring range: Partial oxygen pressure:   0 ... 1100 hPa O  2 Oxygen concentration:  0.0 ... 100.0% O  2 Response time: T  90   <10 s  Accuracy (at 25 ° C, 1013 hPa)  <2 % O  2    ± 0.1% O  2 <25% O  2    ± 0.5% O  2 Scope of supply:  sensor, flow diverter, T-piece